4.2 Article Proceedings Paper

Intravascular micropump for augmented liver perfusion:: First in vivo experience

Journal

ARTIFICIAL ORGANS
Volume 25, Issue 5, Pages 392-394

Publisher

BLACKWELL SCIENCE INC
DOI: 10.1046/j.1525-1594.2001.06772.x

Keywords

rotary blood pump; portal hypertension; mechanical portal pumping

Ask authors/readers for more resources

The aim of this study was to assess the in vivo performance of a new microaxial rotary blood pump developed for long-term intraportal implantation. The pump, measuring 7 mm in diameter, has a single stage impeller and is powered by a microelectric motor. The pump was implanted into the portal vein in 13 large white pigs under general anesthesia. All animals recovered after the portal pump implantation. and they were observed until the pump failed. The 2 longest running pumps performed for 40 and 36 h, respectively. Either thrombus formation or technical problems, especially in the bearings, were the main causes of pump failure during the experiment. No local or systemic adverse effects were observed during the portal pumping period. Full recovery of the animals following intraportal pump implantation was achieved. However, further technical improvements to the pump are required to maintain a longer performance in vivo.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available