4.5 Article Proceedings Paper

Projectile cylinder accidents resulting from the presence of ferromagnetic nitrous oxide or oxygen tanks in the MR suite

Journal

AMERICAN JOURNAL OF ROENTGENOLOGY
Volume 177, Issue 1, Pages 27-30

Publisher

AMER ROENTGEN RAY SOC
DOI: 10.2214/ajr.177.1.1770027

Keywords

-

Ask authors/readers for more resources

Objective. The purpose of this study is to alert MR users to the potentially lethal consequences of projectile cylinder accidents in the MR environment. Conclusion. Projectile cylinder tank accidents still occur and may be increasing, despite adherence to screening policies before MR imaging and safety education of hospital personnel. Four of the last five accidents at our institutions occurred within the past 3 years.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available