ESR dosimetry using inorganic materials:: a case study of Li2CO3 and CaSO4:Dy as prospective dosimeters

Journal

APPLIED RADIATION AND ISOTOPES
Volume 55, Issue 2, Pages 253-258

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/S0969-8043(01)00043-4

Keywords

ESR dosimetry; inorganic materials; TL phosphor

Ask authors/readers for more resources

The COT radical ion, detected by ESR technique in bones and teeth enamel, was proved to be invaluable in high level and retrospective dosimetry. In these matrices, impurity carbonate (at phosphate sites) was the precursor to CO2- With a view to investigate the possibility of using inorganic materials such as lithium carbonate as ESR dosimeters, studies were carried out on gamma-irradiated Li2CO3 The intensity of radiation-induced ESR signals of Li2CO3 at g = 2.0036 (CO3-) and g = 2.0006 (CO2-) was followed as a function of gamma dose in the low dose range of 1-1350 Gy. It was observed that the intensity of the ESR signal at g = 2.0036 (CO3-) was in a linear relation with the radiation done in the dose range 10- 800 Gy and the signal at g = 2.0006 (CO2-) showed linear response in the dose range 5- 800 Gy. The lowest dose that could be detected in the present studies using the signal of CO2- in Li2CO3 powder samples (similar to 50 mg) is 3.2 Gy. ESR studies were also carried out on the widely used TL dosimetric material CaSO4:Dy and in pure CaSO4 after gamma irradiation. The TL materials were used in powder as well as pellet forms. The linearity of ESR response with dose for powder and pellet forms of CaSO4: Dy was also studied using the signals at g = 2.0030 (SO3-) and at g = 2.0139 (SO4-). It was observed that the range of linearity of dose response extended between 20 and 1200 Gy, for SO; signals. The results of dosimetric study indicate that the ESR-Li2CO3 system could be used in dosimetric applications in radiotherapy, However, for the actual applications further advancement is needed to lower the detection limit. The TL phosphor, CaSO4:Dy in powder and pellet forms. could be used as ESR dosimeter in the dose range 20-600 Gy. (C) 2001 Elsevier Science Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available