4.7 Article

Potentiometric Cd2+-selective electrode with a detection limit in the low ppt range

Journal

ANALYTICA CHIMICA ACTA
Volume 440, Issue 2, Pages 71-79

Publisher

ELSEVIER
DOI: 10.1016/S0003-2670(01)01052-2

Keywords

Cd2+-selective electrode; ion-selective electrodes; potentiometry; low detection limits; trace level analysis

Ask authors/readers for more resources

A recently introduced model describing the response curves of polymer membrane ion-selective electrodes (ISEs) is used here to optimize a Cd2+-ISE with a view to achieve very low detection limits. In a first step, the selectivity behavior is determined for two membranes based on the ionophore, N,N,N',N'-tetradodecyl-3,6-dioxaoctanedithioamide (ETH 5435), but having different concentrations of the ionophore and ion-exchanger. Based on these data, the optimal response curves are calculated with a model that takes into account the ion fluxes induced by interfering ions. The experimental results with different ionic backgrounds correspond very well with the predicted effects. The best lower detection limit of 10(-10) M or 11 ppt Cd2+ is achieved at pH 7 with an ionic background of 10(-4) M NaNO3. (C) 2001 Elsevier Science B.V. All lights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available