4.6 Article

Trunnion-Head Stresses in THA: Are Big Heads Trouble?

Journal

JOURNAL OF ARTHROPLASTY
Volume 30, Issue 6, Pages 1085-1088

Publisher

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2015.01.021

Keywords

total hip arthroplasty; Morse taper; trunnion; corrosion; finite element analysis; large diameter head

Categories

Ask authors/readers for more resources

The effects of large heads on stresses at the THA trunnion-head junction and their impact on tribocorrosion/metal ion release remain controversial. A 12/14 3D-model of a stem with different head sizes was investigated. Material properties of titanium were assigned to the trunnion and cobalt-chrome/alumina to the heads. A load simulating walking single-leg stand phase was applied to the head. A total contact head-trunnion interface was assumed. The area underneath the junction underwent significant elevations in stresses as head size increased from 28- to 40-mm. Maximum principal stress doubled between 28 and 40-mm heads, regardless of headmaterial. Stress levels had a direct correlation to head diameter. Stress increases observed using increasingly larger heads will probably contribute to head-trunnion tribocorrosion and ion release. (C) 2015 Elsevier Inc. All rights reserved.