Dual-energy tissue cancellation in mammography with quasi-monochromatic x-rays

Journal

PHYSICS IN MEDICINE AND BIOLOGY
Volume 47, Issue 2, Pages 305-313

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/47/2/309

Keywords

-

Ask authors/readers for more resources

Dual-energy radiography has not evolved into a routine clinical examination yet due to intrinsic limitations of both dual-kVp imaging and single-exposure imaging with conventional x-ray sources. The recent introduction of novel quasi-monochromatic x-ray sources and detectors could lead to interesting improvements, especially in mammography where the complex structure of healthy tissues often masks the detectability of lesions. A dual-energy radiography technique based on a tissue cancellation algorithm has been developed for mammography, with the aim of maximizing the low intrinsic contrast of pathologic tissues while being able to minimize or cancel the contrast between glandular and fat tissues. Several images of a plastic test object containing various tissue equivalent inserts were acquired in the energy range 17-36 keV using a quasi-monochromatic x-ray source and a scintillator-coated CCD detector. Images acquired at high and low energies were non-linearly combined to generate two energy-independent basis images. Suitable linear combinations of these two basis images result in the elimination of the contrast of a given material with respect to another. This makes it possible to selectively cancel certain details in the processed image.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available