Journal
ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 372, Issue 3, Pages 506-510Publisher
SPRINGER HEIDELBERG
DOI: 10.1007/s00216-001-1134-5
Keywords
chromium (III)-selective electrode; sensor; macrocycle
Ask authors/readers for more resources
A new tetradentate dihydrogen perchlorate macrocyclic ligand (2,4,9,11-tetraphenyl-1,5,8,12-tetraazacyclotetradeca-1,4,8,11-tetraene dihydrogen perchlorate) was prepared and characterised. The macrocycle behaves as a selective chelating ion-exchanger for some metal ions. The poly styrene-based membrane electrode is found to exhibit quite promising selectivity for Cr3+ ions. It can be used to estimate chromium concentrations in the range 3.16x 10(-6)-1.00x10(-1) M with a near-Nernstian slope of 17.5 mV per decade of concentration between pH 3.0 to 6.5. The electrode is found to possess a fast response time of 15 s and was used over a period of three months with good reproducibility (s = +/-0.3 mV). The selectivity coefficient values for mono-, di- and trivalent cations indicate excellent selectivity for Cr3+ ions over a large number of other cations. Anions such as Cl- and SO42- do not interfere and the electrode also works satisfactorily in a mixed organic-water solution. The sensor has been used as an indicator electrode for the potentiometric titration of Cr3+ with EDTA. The practical utility of the membrane sensor has also been demonstrated in solutions contaminated with detergents (CTAB and SDS). Above all, the membrane sensor has been very successfully used to determine Cr3+ in some foods.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available