Journal
ANALYTICA CHIMICA ACTA
Volume 452, Issue 2, Pages 223-230Publisher
ELSEVIER SCIENCE BV
DOI: 10.1016/S0003-2670(01)01476-3
Keywords
on-line; preconcentration and separation; FAAS; silver; gold; palladium
Categories
Ask authors/readers for more resources
A simple and highly selective flow injection (FI) on-line preconcentration and separation flame atomic absorption spectrometric (FAAS) method was developed for the determinations of trace amounts of silver, gold and palladium. The selective preconcentration of the noble metals was achieved in a wide range of sample acidity (0.1-6 M HNO3 or HCl) on a microcolumn packed with amidinothioureido-silica gel (ATuSG). The analytes retained on the column were effectively eluted with 5.0% thiourea solution. The analytical procedure was optimized for sample acidity, elution, interferences, flow rate of sampling and eluting, and concentration of sample. Common co-existing cations and anions did not interfere with the preconcentration and determination of the three metals. At a sample loading flow rate of 4.5 ml min(-1) with 60 s preconcentration, detection limits (3sigma) of 1.1 ng ml(-1) Ag, 13 ng ml(-1) Au and 17 ng ml(-1) Pd were obtained. The precisions (R.S.D., n = 11) were 1.2% for Ag, 1.2% for Au and 1.7% for Pd, respectively. The detection limits can be further improved by increasing sample volume. The analytical results obtained by the proposed method for a number of standard reference materials were in good agreement with the certified values. (C) 2002 Elsevier Science B.V. All rights reserved.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available