4.7 Article

Kinetic methods for the determination of cadmium(II) based on a flow-through bulk optode

Journal

TALANTA
Volume 56, Issue 3, Pages 481-489

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/S0039-9140(01)00571-9

Keywords

flow-through optode; kinetic methods; cadmium determination

Ask authors/readers for more resources

Three kinetic methods based on flow injection, flow, and stopped-flow injection were applied for the determination of Cd(II) using a flow-through bulk optode membrane that incorporates 1-(2-pyridylazo)-2-naphtol (PAN) in a plasticized poly(vinyl) chloride membrane entrapped in a cellulose support, The calibration graphs plotting the reflectance at 560 nm versus [Cd(II)](1/2) for the first two methods and versus [Cd(II)] for the third were linear up to 56.2 mg l(-1), The detection limits of the methods were 0.01, 0.06 and 0.8 mg l(-1), respectively. The FI method was selected for application purposes. The variation coefficient of the sensor response for 11.2 mg l(-1) of Cd(II) was +/- 0.31 and +/-0.65% between different membranes. The sensor can be readily regenerated with a carrier (acetic-acetate) buffer of pH 4. The FI method was applied to the determination of cadmium in an alloy and in water. (C) 2002 Elsevier Science B.V, All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available