4.4 Article

Calibration of the electron volt spectrometer, a deep inelastic neutron scattering spectrometer at the ISIS pulsed neutron spallation source

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/S0168-9002(01)00953-6

Keywords

neutron compton scattering; deep inelastic neutron scattering; momentum distributions; electron volt spectrometer; calibration

Ask authors/readers for more resources

The electron Volt Spectrometer (eVS) is an inverse geometry filter difference spectrometer that has been optimised to measure the single atom properties of condensed matter systems using a technique known as Neutron Compton Scattering (NCS) or Deep Inelastic Neutron Scattering (DINS). The spectrometer utilises the high flux of epithermal neutrons that are produced by the ISIS neutron spallation. source enabling the direct measurement of atomic momentum distributions and ground state kinetic energies. In this paper the procedure that is used to calibrate the spectrometer is described. This includes details of the method used to determine detector positions and neutron flight path lengths as well as the determination of the instrument resolution. Examples of measurements on 3 different samples are shown, ZrH2, He-4 and Sn which show the self-consistency of the calibration procedure. (C) 2002 Elsevier Science B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available