Rhodium as permanent modifier for atomization of lead from biological fluids using tungsten filament electrothermal atomic absorption spectrometry

Journal

SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
Volume 57, Issue 4, Pages 727-740

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/S0584-8547(02)00002-2

Keywords

lead (Pb); blood; urine; electrothermal atomic absorption spectrometry; tungsten filament; Rh modifier; interferences

Categories

Ask authors/readers for more resources

Rhodium. (Rh) was investigated as a permanent modifier for the atomization of Pb from biological fluids in W-filament atomic absorption spectrometry (AAS). Heating the W-filament with a Rh solution provided a protective coating for subsequent determinations of Pb in blood and urine matrices. The W-filament AAS instrumentation used was based on a prototype design that utilized self-reversal background correction scheme and peak area measurements. We found that Rh not only stabilized Pb during the pyrolysis step, but also facilitated the removal of carbonaceous residues during the cleaning step, requiring much less power than with phosphate modifier. Thus, the filament lifetime was greatly extended to over 300 firings. Periodic reconditioning with Rh was necessary every 30 firings or so. Conditioning the filament with Rh also permitted direct calibration using simple aqueous Pb standards. The method detection limit for blood Pb was approximately 1.5 mug dl(-1), similar to that reported previously. Potential interferences from concomitants such as Na, K, Ca and Mg were evaluated. Accuracy was verified using lead reference materials from the National Institute of Standards and Technology and the New York State Department of Health. Blood lead results below 40 mug dl(-1) were within +/-1 mug dl(-1) of certified values, and within +/-10% above 40 mug dl(-1): withinrun precision was 10% or better. Additional validation was reported using proficiency test materials and human blood specimens. All blood lead results were within the acceptable limits established by regulatory authorities in the US. When measuring Pb in urine, sensitivity was reduced and matrix-matched calibration became necessary. The method of detection limit was 27 mug l(-1) for urine Pb. Urine lead results were also validated using an acceptable range comparable to that established for blood lead by US regulatory agencies. (C) 2002 Elsevier Science B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available