4.8 Article

Thermal vaporization-vacuum ultraviolet laser ionization time-of-flight mass spectrometry of single aerosol particles

Journal

ANALYTICAL CHEMISTRY
Volume 74, Issue 9, Pages 2048-2052

Publisher

AMER CHEMICAL SOC
DOI: 10.1021/ac011225a

Keywords

-

Ask authors/readers for more resources

Single aerosol particles of ethylene glycol and oleic acid are vaporized on a heater at temperatures between 500 and 700 1, and the resulting vapor plume is ionized by a 10.5-eV vacuum ultraviolet (VUV) laser. The mass spectra are compared to those obtained by CO2 laser vaporization followed by VUV laser ionization. The relative intensities of the parent and fragment ion peaks are remarkably similar for the two modes of vaporization. A Maxwell-Boltzmann distribution of speeds accurately describes the dependence of the signal as a function of the VUV laser pulse timing. The signal levels obtained with this design are sufficient to obtain good-quality mass spectra.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available