4.7 Article

Electrical aerosol spectrometer of Tartu University

Journal

ATMOSPHERIC RESEARCH
Volume 62, Issue 3-4, Pages 315-324

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/S0169-8095(02)00017-0

Keywords

aerosol number size distribution; parallel measurement principle; high time resolution; real time data inversion

Ask authors/readers for more resources

The electrical aerosol spectrometer (EAS) of the parallel measuring principle at Tartu University is an efficient instrument for rapid measurement of the unstable size spectrum of aerosol particles. The measuring range from 10 tun to 10 mum is achieved by simultaneously using a pair of differential mobility analyzers with two different particle chargers. The particle spectrum is calculated and measurement errors are estimated in real time by using a least-squares method. Experimental calibration ensures reliability of measurement. The instrument is well suited for continuous monitoring of atmospheric aerosol. (C) 2002 Elsevier Science B.V. All rights reserved.