4.8 Article

Fracture mechanisms of retrieved titanium screw thread in dental implant

Journal

BIOMATERIALS
Volume 23, Issue 12, Pages 2459-2465

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/S0142-9612(01)00380-5

Keywords

titanium; dental implant; delayed fracture; metallurgical structure; environmental corrosion; shear crack; hydrogen

Ask authors/readers for more resources

Titanium and its alloy are increasingly attracting attention for use as biomaterials. However, delayed fracture of titanium dental implants has been reported, and factors affecting the acceleration of corrosion and fatigue have to be determined. The fractured surface of a retrieved titanium screw and metallurgical structures of a dental implant system were analyzed. The outer surface of the retrieved screw had a structure different from that of the as-received screw. It was confirmed that a shear crack initiated at the root of the thread and propagated into the inner section of the screw. Gas chromatography revealed that the retrieved screw had absorbed a higher amount of hydrogen than the as-received sample. The grain structure of a titanium screw, immersed in a solution known to induce hydrogen absorption, showed features similar to those of the retrieved screw. It was concluded that titanium in a biological environment absorbs hydrogen and this may be the reason for delayed fracture of a titanium implant. (C) 2002 Published by Elsevier Science Ltd.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available