4.5 Article

The effect of preflaring on the rates of separation for 0.04 taper nickel titanium rotary instruments

Journal

JOURNAL OF ENDODONTICS
Volume 28, Issue 7, Pages 543-545

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/00004770-200207000-00015

Keywords

-

Ask authors/readers for more resources

The purpose of this study was to compare the rates of separation of 0.04 taper nickel titanium (NiTi) rotary instruments using two different instrumentation techniques. Twenty sets of 0.04 taper Profile Series 29 rotary instruments, sizes 2 to 6 were used in the mesial (mandibular) or buccal (maxillary) canals of extracted human molars with a 20 to 30 degree root curvature according to the Schneider classification. The rotary instruments were used up to 20 times either with the crown-down technique recommended by the manufacturer or with a combination of preflaring with hand files in a passive step-back technique followed by rotary instrumentation. Statistical analysis of the data showed that the combination technique allowed more uses before separation compared with the crown-down technique recommended by the manufacturer p < 0.0001.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available