4.7 Article

The effects of automatic artifact rejection on evoked potential recordings

Journal

COMPUTERS IN BIOLOGY AND MEDICINE
Volume 32, Issue 4, Pages 247-259

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/S0010-4825(02)00019-7

Keywords

evoked potential; artifact; noise; intra-operative monitoring; statistics

Ask authors/readers for more resources

An analysis of the effects of automatic artifact rejection on the evoked potential recordings is made using both a theoretical analysis and computer simulations. It is demonstrated that when the noise at each measurement point is independently distributed with maximum probability near zero amplitude, the bias and changes in variance induced by the automatic artifact rejection are small unless more than 80% of traces are rejected. When high-amplitude noise transients occur, there can be significant bias in the evoked potential when the artifact rejection threshold is set near the amplitude of the noise transients. Even more substantial distortions in the evoked potential occur when the noise is not independently distributed at each measurement point. It is important to recognize the possibility of these effects during clinical evoked potential recording. (C) 2002 Elsevier Science Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available