The effect of environmental conditions and electrical charge on the weighing accuracy of different filter materials

Journal

SCIENCE OF THE TOTAL ENVIRONMENT
Volume 293, Issue 1-3, Pages 201-206

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/S0048-9697(02)00015-3

Keywords

filters; environmental conditions; electrical charge; weighing accuracy

Ask authors/readers for more resources

Different filter materials and electrical charge elimination methods were used to investigate the weighing accuracy of filter papers under different environmental conditions. The results show that the standard deviations (S.D.) of weight data for glass fiber and MCE filters were substantial without environmental control, whether or not the electrical charge eliminators were used. Values of 0.157 and 0.349 mg were determined for glass and NICE filters, respectively. The accuracy of weighing was substantially improved and the S.D. was reduced to 0.01 and 0.09 mg for glass fiber and MCE filters, respectively, after applying the environmental control conditions. For PVC and Teflon filters, the accuracy of weighing was good, even in the uncontrolled environmental conditions, whether or not the electrical charge eliminators were used. The S.D. values of weighing data of PVC and Teflon filters were 0.007 and 0.011 mg, respectively. (C) 2002 Elsevier Science B.V. All rights reserved.