4.7 Article

Glucose biosensors prepared by electropolymerization of p-chlorophenylamine with and without Nafion

Journal

ANALYTICA CHIMICA ACTA
Volume 463, Issue 2, Pages 239-247

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/S0003-2670(02)00430-0

Keywords

glucose oxidase; biosensor; nafion; poly(p-chlorophenylamide)

Ask authors/readers for more resources

Two different glucose biosensors for the amperometric determination of glucose, based on poly (p-chlorophenylamide) (PCPA) and bilayer film of PCPA and Nafion (PCPA/Nafion), are successfully developed. These two biosensors show linear amperometric responses to glucose ranging from 2.0 x 10(-4) to 3.5 x 10(-2) mol(-1) and 5.0 x 10(-4) to 7.5 x 10(-2) mol l(-1), respectively, with the same correlation coefficient of 0.9988, Effects of polymerization potential and polymerization time on the performance of enzyme sensors are studied, It is found that PCPA, as a non-conducting polymer, can largely reduce the influence of electroactive interferents. Introduction of inner Nafion membrane not only further eliminates the influence of ascorbic acid on the sensor response but also increases electrode stability. (C) 2002 Elsevier Science B.V. All rights reserved.