4.1 Article

Heat generation during implant drilling: The significance of motor speed

Journal

JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY
Volume 60, Issue 10, Pages 1160-1169

Publisher

W B SAUNDERS CO-ELSEVIER INC
DOI: 10.1053/joms.2002.34992

Keywords

-

Ask authors/readers for more resources

Purpose: The purpose of this study was to measure the heat generated from 3 drilling speeds (1,225, 1,667, and 2,500 rpm) using the armamentarium of 4 implant systems. Materials and Methods: The mean rise in temperature, the time of drilling, and the time needed for pig jaw bone to return to the baseline temperature were monitored using 4 thermocouple technology. Results: The mean rise in temperature, the time of drilling, and the time needed for the specimens to return to the baseline temperature were lower at 2,500 rpm than at 1,667 or 1,225 rpm (P less than or equal to .05), regardless of the system used. The rpm also directly correlated to the amount of time the bone remained at an elevated temperature. Conclusion: From a heat generation standpoint, we conclude that preparing an implant site at 2500 rpm could decrease the risk of osseous damage, which may affect the initial healing of dental implants. This may decrease the devital zone adjacent to an implant after surgery and be most advantageous in immediate load application to dental implants. (C) 2002 American Association of Oral and Maxillo/acial Surgeons.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available