4.7 Article

Membrane-immobilized haptoglobin as affinity matrix for a hemoglobin-A1c immunosensor

Journal

ANALYTICA CHIMICA ACTA
Volume 470, Issue 2, Pages 111-119

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/S0003-2670(02)00779-1

Keywords

HbA1c; glycated hemoglobin; diabetes; immunosensor; haptoglobin

Ask authors/readers for more resources

An amperometric immunosensor for hemoglobin-A1c (HbA1c) determination has been developed utilizing membrane-immobilized haptoglobin as affinity matrix fixed in front of a Pt-working electrode. The HbA1c assay was carried out in a two-step procedure including the selective hemoglobin enrichment on the sensor surface and the specific HbA1c detection by a glucose oxidase (GOx) labeled anti-HbA1c antibody. Hydrogen peroxide generated by the enzyme label was oxidized at +600 mV versus Ag/AgCl. A standard curve for HbA1c was obtained with a linear range between 0 and 25% HbA1c of total hemoglobin which correspond to 7.8-39 nM. ELISA studies confirmed the advantage of a sandwich-type format with haptoglobin as capture molecule for selective hemoglobin binding over the direct adsorption method. Results by the sandwich immunoassay showed a linear correlation within the clinically relevant range 5-20% (CV < 3). For sensor application the immobilization procedure of haptoglobin onto CDI-activated cellulose membranes was optimized. (C) 2002 Elsevier Science B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available