Cytokine dialysis:: An ex vivo study

Journal

ASAIO JOURNAL
Volume 48, Issue 6, Pages 650-653

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/00002480-200211000-00013

Keywords

-

Ask authors/readers for more resources

To test the hypothesis that dialysis using a new large pore membrane would achieve effective cytokine removal, blood from six volunteers was incubated with endotoxin (1 mg) and then circulated through a closed circuit with a polyamide membrane (nominal cut-off: 100 kDa). Hemodialysis was conducted at 1 or 9 L/hr of dialysate flow at the start of circulation and after 2 and 4 hours. The peak dialysate/plasma concentration ratios were 0.92 for interleukin (IL-1beta, 0.67 for IL-6, 0.94 for IL-8, 0.33 for tumor necrosis factor (TNF)-alpha, and 0.11 for albumin. The dialysate/plasma ratios for all cytokines and albumin were decreased with increased dialysate flow from I to 9 L/hr (p < 0.05). Clearances for IL-1beta, IL-6, and IL-8, however, were significantly improved with increased dialysate flow (p < 0.01). There was no increase in TNF-alpha clearance (not significant) and a decrease in albumin clearance (p < 0.01). The peak clearance at 9 L/hr was 33 ml/min for IL-1beta, 19 for IL-6, 51 for IL-8, 11 for TNF-alpha, and 1.2 for albumin. No adsorption of cytokines was observed. We conclude that cytokine dialysis is achievable through a membrane with a high cut-off point with negligible albumin loss. These findings support the technical feasibility of this new approach to blood purification in sepsis.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available