4.6 Article

Determination of potassium ions in pharmaceutical samples by FIA using a potentiometric electrode based on ionophore nonactin occluded in EVA membrane

Journal

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/S0731-7085(02)00598-8

Keywords

potassium determination; potentiometric electrode; pharmaceutical samples

Ask authors/readers for more resources

A simple and rapid method was developed for the K+ ions determination employing a flow injection system using a flow-through electrode based on the naturally-occurring antibiotic ionophore nonactin occluded in a polymeric membrane. The nonactin ionophore was trapped in poly(ethylene-co-vinyl acetate) (EVA) matrix (40% w/w in vinyl acetate) and dispersed on the surface of a graphite-epoxy tubular electrode. The plasticizer-free all-solid-state potassium-selective electrode showed a linear response for K+ concentrations between 5.0 x 10(-5) and 5.0 x 10(-2) M (r = 0.9995) with a near-Nernstian slope of 51.5 mV per decade, when Tris-HCI buffer (pH 7.0;0,1 M) was employed as a carrier. The potentiometric-FIA system allows an analytical frequency of 120 samples per hour with a precision of 3.6%. The relative standard deviations (R.S.D.) for K+ determination in pharmaceuticals samples, without any previous treatment, were lower than 4.0%, comparable to those obtained by flame photometry. Ammonium is the main analytical interference and the electrode response time was 5 s at 25 degreesC. The useful lifetime of the tubular sensor is longer than 3 months in continuous use. (C) 2002 Elsevier Science B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available