4.7 Article

On-line determination of trace levels of palladium by flame atomic absorption spectrometry

Journal

TALANTA
Volume 59, Issue 4, Pages 651-657

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/S0039-9140(02)00616-1

Keywords

Pd(II)preconcentration; on-line sub-mg 1(-1) determination; flame atomic absorption spectrometry; metalfix-chelamine resin

Ask authors/readers for more resources

A preconcentration method is developed for the on-line determination of palladium in complex matrices with flame atomic absorption spectrometry (FAAS). The flow system comprised of a minicolumn filled with polyamine Metalfix-Chelamine resin which is highly selective for Pt(IV), Au(III) and Pd(II). Best preconcentration conditions are established by testing different resin quantities, sample and eluent solution volumes, and adsorption and elution steps flow rates. Sample volumes of 4.7 ml of palladium solutions resulted in an enrichment factor of twenty at the optimum hydrodynamic conditions. This value can be increased by injecting larger volumes of sample solution. The method is sensitive, easy to operate and permitted the determination of sub-mg 1(-1) levels of palladium with a detection limit of 0.009 mg 1(-1). The resin was used up to 60 times in consecutive retention-elution cycles without any appreciable deterioration in its performance. The applicability of this method was tested by determining the palladium content in synthetic geological samples as well as in the pellet-type used car catalyst reference material. (C) 2002 Elsevier Science B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available