4.8 Article Proceedings Paper

Microdevice with integrated dialysis probe and biosensor array for continuous multi-analyte monitoring

Journal

BIOSENSORS & BIOELECTRONICS
Volume 18, Issue 5-6, Pages 613-619

Publisher

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/S0956-5663(03)00038-1

Keywords

BioMEMS; microdialysis; glucose; lactate; continuous monitoring; microfabrication

Ask authors/readers for more resources

The simultaneous on-line determination of glucose and lactate using a microdevice that consisted of a dialysis sampling system incorporated to the flow-through cell of a microfabricated biosensor array is presented. The fluidic connections between the different device's components were realized by subsequent processing of stacked dry resist layers on a plastic support that provided also the means for electric connections. The performance of the device was evaluated in vitro. The cross-talk effect on the downstream sensor was investigated and found to be negligible. Recoveries of over 95% for both analytes were achieved when flow rates of the perfusion fluid less than or equal to 0.5 mul/min were used. At this flow rate, the response time of the device was 2.4 min, which is acceptable for on-line analysis. The linear response concentration range extended up to 30 mM for glucose and 15 mM for lactate. Interference from electroactive species such as ascorbic acid, 2-acetamidophenol and uric acid, was minimal (less than 5% increase in biosensors signal for all substances tested). In addition, the device presented long-term run stability both in buffer and serum samples. (C) 2003 Elsevier Science B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available