Volume 30, Issue 5, Pages 986-992Publisher
AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.1567736
Keywords
Gamma Knife; 4 mm helmet; radiation detectors
Ask authors/readers for more resources
Though the 4 mm Gamma Knife helmet is used routinely, there is disagreement in the Gamma Knife users community on the value of the 4 mm helmet relative output factor. A range of relative output factors is used, and this variation may impair observations of dose response and optimization of prescribed dose. To study this variation, measurements were performed using the following radiation detectors: silicon diode, diamond detector, radiographic film, radiochromic film, and TLD cubes. To facilitate positioning of the silicon diode and diamond detector, a three-dimensional translation micrometer was used to iteratively determine the position of maximum detector response. Positioning of the films and TLDs was accomplished by manufacturing custom holders for each technique. Results from all five measurement techniques indicate that the 4 mm helmet relative output factor is 0.868 +/- 0.014. Within the experimental uncertainties, this value is in good agreement with results obtained by other investigators using diverse techniques. (C) 2003 American Association of Physicists in Medicine.