4.0 Article

Electroanalytical determination of 4-nitrophenol by square wave voltammetry on diamond electrodes

Journal

JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY
Volume 14, Issue 4, Pages 530-535

Publisher

SOC BRASILEIRA QUIMICA
DOI: 10.1590/s0103-50532003000400007

Keywords

4-nitrophenol; square wave voltammetry; diamond electrodes

Ask authors/readers for more resources

The anodic voltammetric behavior of 4-Nitrophenol on a Boron-doped diamond electrode in aqueous solution has been studied using square wave voltammetry. After optimization of the experimental conditions, that model molecule was analyzed in pure water solutions using a Britton-Robinson buffer with pH 6.0 as the supporting electrolyte. Oxidation occurs at 1.0 V vs. Ag/AgCl in an irreversible two-electron process controlled by adsorption of the species. The detection limit (DL) obtained was 2.8 mug L-1. This result was comparable to that obtained from reduction of the molecule at - 0.8 V vs. Ag/AgCl under the same experimental conditions (DL = 4.2 mug L-1). Both DL values are within the limits required by the legislation for drinking water (30 mug L-1). The combination of square wave voltammetry and diamond electrodes has proved to be an interesting and desirable alternative for the analytical determination of organic molecules.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.0
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available