4.4 Article

Choice of data-collection parameters based on statistic modelling

Journal

ACTA CRYSTALLOGRAPHICA SECTION D-STRUCTURAL BIOLOGY
Volume 59, Issue -, Pages 1145-1153

Publisher

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S0907444903008163

Keywords

-

Ask authors/readers for more resources

A new method and the software program BEST for optimal planning of X-ray data collection from protein crystals using the rotation method are presented. From one or a few initial diffraction images, BEST estimates the statistical characteristics of the data set for different combinations of data-collection parameters and suggests the most optimal ones The. anisotropy in diffraction and the permitted width of oscillation without spatially overlapping reflections are taken into account. According to the option chosen, the optimal set of parameters provides a given average signal-to-noise ratio at a given resolution either in the shortest time or with the minimum total radiation dose. BEST has been successfully used at the protein crystallography beamlines at DORIS (DESY). The software proved to be extremely useful in using the available data-collection time in the most efficient way.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available