Purpose: To compare three-dimensional (3D) spatial-spectral (SS) spoiled gradient-recalled acquisition in the steady state (SPGR) imaging with fat-suppressed 3D SPGR sequences in MR imaging of articular cartilage of the knee joint in patients with osteoarthritis. Materials and Methods: MR images of six patients with osteoarthritis of the knee were prospectively examined with a 1.5T MR scanner. For quantitative analyses, the signal-to-noise ratios, contrast-to-noise ratios, and contrast of cartilage and adjacent structures including meniscus, synovial fluid, muscle, fat tissue, and bone marrow were measured. Results: In patients with osteoarthritis, 3DSS-SPGR images demonstrated higher spatial resolution and higher mean signal-to-noise (S/N) ratios (cartilage, 24.9; synovial fluid, 12.3; muscle, 20.7; meniscus, 21.6), with shorter acquisition times (7 minutes 20 seconds), when compared to fat-suppressed 3D SPGR images (cartilage, 22.3; synovial fluid, 10.8; muscle, 16.7; meniscus, 13.4). Conclusion: 3DSS-SPGR imaging is a promising method for evaluating cartilage pathology in patients with osteoarthritis of the knee and has the potential to replace fat-suppressed 3D SPGR imaging.