4.7 Article

Fabrication of a disposable biosensor for Escherichia coli O157:H7 detection

Journal

IEEE SENSORS JOURNAL
Volume 3, Issue 4, Pages 345-351

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JSEN.2003.815782

Keywords

conductimetric; electrochemical immunoassay; immunosensor; polyaniline

Ask authors/readers for more resources

As the safety in the food supply becomes critical, the demand for a rapid, low-volume, and sensitive microbial detection device has dramatically increased. A biosensor based on an electrochemical sandwich immunoassay using polyaniline has been developed for detecting foodborne pathogens, such as Escherichia coli (E. coli) O157:H7. The biosensor is comprised of two types of proteins: capture protein and reporter protein. The capture protein is immobilized on a pad between two electrodes, while the reporter protein is attached to conductive polymers. After adding the sample, the target protein binds to the reporter protein and forms a sandwich complex with the capture protein. The conductive polymer that is attached to the reporter protein serves as a messenger, reporting the amount of target protein captured in the form of an electrical signal. The architecture of the biosensor utilizes lateral flow format, which allows the liquid sample to move from one pad to another by capillary action. Experiments to evaluate the best construction materials, the optimal polyaniline and antibody concentrations, and the distance between electrodes are highlighted in this paper. Results show that the biosensor could detect approximately 7.8 x 10(1) colony forming unit per mililiter of E. coli O157:H7 in 10 min.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available