4.4 Article

Intraaneurysmal flow changes affected by clip location and occlusion magnitude in a lateral aneurysm model

Journal

MEDICAL ENGINEERING & PHYSICS
Volume 25, Issue 7, Pages 581-589

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/S1350-4533(03)00078-X

Keywords

hemodynamics; lateral aneurysm; embolization; surgical clip; particle image velocitimeter

Ask authors/readers for more resources

In order to study the flow dynamic changes inside an aneurysm sac due to the partial occlusion of the aneurysm neck, velocity fields were measured using a particle image velocitimeter (PIV) in in vitro aneurysm models under the physiological flow waveform. Lateral aneurysm models arising from the curved parent vessel with different occlusion ratios and sites-e.g. no clip, 50% proximal and distal clip, and 75% proximal and distal clip-were tested. Reduced inflow and intraaneurysmal velocities may provide a better hemodynamic environment for aneurysm embolization. Comparing inflow rates and averaged intraaneurysmal velocities in the proximal and the distal clip model, they were lower in the distal clip model in cases of 50% neck occlusion, but they were lower in the proximal clip model in cases of 75% occlusion. These results suggest that clipping sites for reduced inflow and intraaneurysmal flow velocities may differ for different residual neck sizes. Less effective inflow blocking in the 75% distally clipped model may be due to the curvature of the parent artery. Therefore, not only the residual neck size and clipping site but the geometry of parent vessel significantly affect the flow fields inside the aneurysm, and subsequently the success of the aneurysm treatment. (C) 2003 IPEM. Published by Elsevier Science Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available