Evaluation of the gamma dose distribution comparison method

Journal

MEDICAL PHYSICS
Volume 30, Issue 9, Pages 2455-2464

Publisher

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.1598711

Keywords

radiation therapy; intensity modulated radiation therapy; dose distribution comparison; gamma tool

Funding

  1. PHS HHS [R01 84409] Funding Source: Medline

Ask authors/readers for more resources

The gamma tool was developed to quantitatively compare dose distributions, either measured or calculated. Before computing gamma, the dose and distance scales of the two distributions, referred to as evaluated and reference, are renormalized by dose and distance criteria, respectively. The renormalization allows the dose distribution comparison to be conducted simultaneously along dose and distance axes. The gamma quantity, calculated independently for each reference point, is the minimum distance in the renormalized multidimensional space between the evaluated distribution and the reference point. The gamma quantity degenerates to the dose-difference and distance-to-agreement tests in shallow and very steep dose gradient regions, respectively. Since being introduced, the gamma quantity has been used by investigators to evaluate dose calculation algorithms, and compare dosimetry measurements. This manuscript examines the gamma distribution behavior in two dimensions and evaluates the gamma distribution in the presence of data noise. Noise in the evaluated distribution causes the gamma distribution to be underestimated relative to the no-noise condition. Noise in the reference distribution adds noise in the gamma distribution in proportion to the normalized dose noise. In typical clinical use, the fraction of points that exceed 3% and 3 mm can be extensive, so we typically use 5% and 2-3 mm in clinical evaluations. For clinical cases, the calculation time is typically 5 minutes for a 1 X 1 mm(2) interpolated resolution on an 800 MHz Pentium 4 for a 14.1 X 15.2 cm(2) radiographic film. (C) 2003 American Association of Physicists in Medicine.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available