4.5 Article

Electromagnetic field-based image-guided spine surgery part two: Results of a cadaveric study evaluating thoracic pedicle screw placement

Journal

SPINE
Volume 28, Issue 17, Pages E351-E354

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/01.BRS.0000086822.76638.76

Keywords

image guided; thoracic spine; fluoroscopy; pedicle screw; electromagnetic field

Ask authors/readers for more resources

Study Design. Human cadaveric. Objectives. Compare the accuracy of electromagnetic field (EMF)-based image-guided thoracic pedicle screw insertion to conventional techniques using anatomic landmarks and fluoroscopy. Background. Image-guided surgical systems that aid in spinal instrumentation seek to minimize radiation exposure and improve accuracy. EMF image guidance was developed as an alternative to optical tracking to eliminate potential line of sight issues. Materials and Methods. Four fresh-frozen human cadavers were randomly allocated into two groups. Pedicle screws were inserted from T1 to T12 using anatomic landmarks and fluoroscopy in group 1 and image guidance in group 2. Insertion and fluoroscopy time were recorded. Anatomic dissections were performed to assess screw placement. Results. Image guidance placed 92% of thoracic pedicle screws safely, and conventional fluoroscopy placed 90% safely. The average degree of perforation was 2.4 mm with conventional fluoroscopy and 1.7 mm with image guidance (P = 0.055). Fluoroscopic time per screw was 5.9 seconds for conventional fluoroscopy and 3.6 seconds for image guidance (P = 0.045). Insertion time per screw was 4.35 minutes for conventional fluoroscopy and 2.98 minutes for image guidance (P = 0.007). However, when set-up time and image capture time were taken into account, the total insertion time per screw was not significantly different between the two groups. Conclusions. Our study has shown that EMF image-guided thoracic pedicle screw placement results in a similar incidence of safely placed screws as does conventional fluoroscopy. When set-up time and image-capture time were factored in for image guidance, the average time to insert a pedicle screw was equal for both techniques. The use of EMF image guidance significantly reduced fluoroscopic time and thus radiation exposure per screw compared with conventional fluoroscopic techniques.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available