4.6 Article

Optimization of a microphone for photoacoustic spectroscopy

Journal

APPLIED SPECTROSCOPY
Volume 57, Issue 9, Pages 1087-1092

Publisher

SAGE PUBLICATIONS INC
DOI: 10.1366/00037020360695946

Keywords

photoacoustie spectroscopy; optimization; microphone; nonresonant; photoacoustic FT-IR spectroscopy

Ask authors/readers for more resources

The principles of optimizing a microphone for a photoacoustic detector in the nonresonant mode of operation are presented. Performance of the constructed optical cantilever microphone is compared to the commercial electret microphone by measuring the photoacoustic spectrum of methane with both systems. The sensitivity of the presented photoacoustic detector in the nonresonant mode of operation is 100 times higher with the optical cantilever microphone than with the electret microphone.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available