4.8 Article

Cl- sensitive biosensor used electrolyte-solution-gate diamond FETs

Journal

BIOSENSORS & BIOELECTRONICS
Volume 19, Issue 2, Pages 137-140

Publisher

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/S0956-5663(03)00174-X

Keywords

hydrogen-terminated diamond; oxygen-terminated diamond; Cl- and Br- ionic solution; threshold voltage; cystic fibrosis

Ask authors/readers for more resources

We have investigated the electrolyte-solution-gate field effect transisitors (SGFETs) used hydrogen terminated (H-terminated) or partially oxygen terminated (O-terminated) polycrystalline diamond surface in the Cl- and Br- ionic solutions. The H-terminated channel SGFETs are insensitive to pH values in electrolyte solutions. The threshold voltages of the diamond SGFETs shift according to the density of Cl(-)and Br- ions about 30 mV/decade. One of the attractive biomedical applications for the Cl- sensitive SGFETs is the detection of chloride density in blood or in sweat especially in the case of cystic fibrosis. The sensitivities of Cl- and Br- ions have been lost on the partially O-terminated diamond surface. These phenomena can be explained by the polarity of surface change on the H-terminated and the O-terminated surface. (C) 2003 Elsevier B.V. All rights reserved.