4.4 Article

Impact of strut thickness on late luminal loss after coronary artery stent placement

Journal

AMERICAN JOURNAL OF CARDIOLOGY
Volume 93, Issue 4, Pages 477-480

Publisher

EXCERPTA MEDICA INC-ELSEVIER SCIENCE INC
DOI: 10.1016/j.amjcard.2003.10.049

Keywords

-

Ask authors/readers for more resources

To determine the influence of coronary artery stent strut thickness on angiographic late luminal loss, 663 patients were included in a single-center observational cohort after receiving an ACS Multilink stent in a native coronary vessel. At 6- to 10-month follow-up, 287 patients treated with a thin-strut stent (50 mum) had significantly less late luminal loss than 376 patients treated with a thick-strut stent (greater than or equal to90 mum) (mean 0.92 +/- 0.59 vs 1.06 +/- 0.71 mm, p = 0.011); on multivariate regression analysis, strut thickness was found to be an independent predictor for late luminal loss. (C) 2004 by Excerpta Medica, Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available