NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
Volume 520, Issue 1-3, Pages 348-350Publisher
ELSEVIER SCIENCE BV
DOI: 10.1016/j.nima.2003.11.264
Keywords
transition edge sensor; reduced excess noise; calorimeter; molydenum; gold; bilayer
Ask authors/readers for more resources
We measured noise in a variety of Mo/Au transition-edge sensor (TES) X-ray calorimeters. We investigated the relationship between the noise, bias, and the superconducting phase transition in the TESs. Our square TES calorimeters have achieved very good energy resolutions (2.4 eV at 1.5 keV) but their resolutions have been limited by broadband white excess noise generated by the TES when it is biased in the phase transition. We have recently fabricated Mo/Cu TESs with interdigitated normal metal bars deposited on top of the bilayer. The new TES calorimeters have demonstrated little or no excess noise in the phase transition. These results point the way to development of TES calorimeters with higher energy resolution. (C) 2003 Elsevier B.V. All rights reserved.