4.5 Article

Evaluation of reversed-phase columns for the analysis of heterocyclic aromatic amines by liquid chromatography -: electrospray mass spectrometry

Publisher

ELSEVIER
DOI: 10.1016/j.jchromb.2003.09.023

Keywords

food analysis; heterocyclic aromatic amines

Ask authors/readers for more resources

Liquid chromatography coupled to mass spectrometry (LC-MS), especially by the use of electrospray ionisation source (ESI), is currently used for the analysis of heterocyclic aromatic amines (HAs) in complex samples. The present paper describes the study of the performance of different narrow-bore reversed-phase columns to achieve the best chromatographic separation for the determination of 16 HAs by LC-ESI-MS in food samples. Different parameters such as peak symmetry, resolution and number of theoretical plates have been evaluated for each column, using different chromatographic conditions. The column that provided the best results was TSK Gel(R) Semi-Micro ODS-80TS of Tosohaas. Quality parameters have been established, obtaining good short-term precision in all cases (relative standard deviation (R.S.D.) lower than 7.7%) and low limits of detection (< 13 pg injected in MS and < 16 pg injected in MS/MS). The content of HAs in two beef extracts have been determined. (C) 2003 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available