4.5 Article

Implant design and interface force transfer - A photoelastic and strain-gauge analysis

Journal

CLINICAL ORAL IMPLANTS RESEARCH
Volume 15, Issue 2, Pages 249-257

Publisher

WILEY
DOI: 10.1111/j.1600-0501.2004.00979.x

Keywords

biomechanics; implant design; oral implants; photoelastic stress analysis; strain-gauge analysis

Ask authors/readers for more resources

Objectives: To compare stress and strain magnitudes of butt-joint and internal-cone oral implants in a bone simulant. Material and methods: Photoelastic and strain-gauged models of solitary Branemark((R)), ITI(R) and Astra Tech((R)) implants were obtained. Vertical and 20degrees oblique forces of 100 and 150 N were applied on the abutments in separate load cases. Isochromatic fringe patterns around implants were observed and photographed in the field of a polariscope. Strain-gauge measurements were performed and principal strains induced around implants were calculated. Results: Isochromatic fringe orders of all designs were similar under both loading conditions (P>0.05). Strains around Branemark((R)) implants were lower than around Astra Tech((R)) and ITI(R) implants, particularly under vertical loads. The latter implants had similar strain magnitudes (P>0.05), although strains around the conical Astra Tech((R)) implant were lower under vertical load. Conclusions: Butt-joint and internal-cone oral implants have similar force distribution characteristics. The implant-abutment mating design is not a decisive factor affecting stress and strain magnitudes in a bone simulant.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available