4.6 Article

Application of off-line size-exclusion chromatographic fractionation-matrix assisted laser desorption ionization time of flight mass spectrometry for proanthocyanidin characterization

Journal

JOURNAL OF CHROMATOGRAPHY A
Volume 1033, Issue 2, Pages 291-297

Publisher

ELSEVIER
DOI: 10.1016/j.chroma.2004.01.052

Keywords

calibration curves; size-exclusion chromatography; matrix-assisted laser desorption ionization mass spectrometry; molecular mass distribution; procyanidins

Ask authors/readers for more resources

In this work. we wanted to devise a reliable method to characterize polymerized forms of tannins, their structural information and mass distribution. Size-exclusion chromatography (SEC) is a chromatographic technique used to determine molecular mass (weight) distributions of polymers. One important step in the data treatment is the modeling of the calibration curves. Polystyrenes (PS) are standards usually used because no commercial procyanidin (PC) standards are available. An off-line coupling of SEC and MALDI was carried out to measure differences between polystyrenes and procyanidins. Thus, a new calibration curve was established; from 1000 to 8000 Da, there is a good correlation between the MALDI and PS calibration curves, in this field the PS calibration is correct and enables true mass determination. For masses above 8000 Da. PS calibration overestimates the real molecular weight of PC, overestimation of 53%. And for masses below 1000 Da, PS calibration underestimates their real molecular weight (10-15%). This means that to truly characterize PC, calibration based on PC standards is required. (C) 2004 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available