4.7 Article

Determination of aluminum(III) in crystallized fruit samples using a multicommutated flow system

Journal

JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY
Volume 52, Issue 9, Pages 2450-2454

Publisher

AMER CHEMICAL SOC
DOI: 10.1021/jf049812v

Keywords

multicommutation; zone sampling; aluminum(III); crystallized fruits; spectrophotometry

Ask authors/readers for more resources

A multicommutated flow system was developed for the determination of aluminum in crystallized fruit samples. Spectro photo metric determination is based on the reaction of aluminum with chrome azurol S. The binary sampling technique was implemented to improve mixing conditions and to minimize reagent consumption. Three different working zones were established (0.5-5.0, 5.0-25.0, and 10.0-100 ppm) using the zone sampling approach, allowing us to adapt the extent of the in-line dilution. The influence of the chemical and physical parameters on the performance of the system was studied. Detection limits of 0.1, 0.6, and 0.8 ppm were obtained for the lowest, the medium, and the highest dispersion system, respectively. The procedure was applied to the determination of aluminum in crystallized fruit extracts. The results were in agreement with those obtained by the reference flame atomic absorption procedure at a 95% confidence level. Repeatability (RSD) was better than 2.4% in all of the three application zones.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available