4.5 Article

Removal of EPI Nyquist ghost artifacts with two-dimensional phase correction

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 51, Issue 6, Pages 1247-1253

Publisher

WILEY
DOI: 10.1002/mrm.20097

Keywords

echo planar imaging; EPI; Nyquist ghost artifact

Funding

  1. NIMH NIH HHS [R01 MH58912] Funding Source: Medline

Ask authors/readers for more resources

Odd-even echo inconsistencies result in Nyquist ghost artifacts in the reconstructed EPI images. The ghost artifacts reduce the image signal-to-noise ratio and make it difficult to correctly interpret the EPI data. In this article a new 2D phase mapping protocol and a postprocessing algorithm are presented for an effective Nyquist ghost artifacts removal. After an appropriate k-space data regrouping, a 2D map accurately encoding low- and high-order phase errors is derived from two phase-encoded reference scans, which were originally proposed by Hu and Le (Magn Reson Med 36:166-171;1996) for their 1D nonlinear correction method. The measured phase map can be used in the postprocessing algorithm developed to remove ghost artifacts in subsequent EPI experiments. Experimental results from phantom, animal, and human studies suggest that the new technique is more effective than previously reported methods and has a better tolerance to signal intensity differences between reference and actual EPI scans. The proposed method may potentially be applied to repeated EPI measurements without subject movements, such as functional MRI and diffusion coefficient mapping. (C) 2004 Wiley-Liss, Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available