4.5 Article

Stainless and shape memory alloy coronary stents: a computational study on the interaction with the vascular wall

Journal

BIOMECHANICS AND MODELING IN MECHANOBIOLOGY
Volume 2, Issue 4, Pages 205-217

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s10237-004-0039-6

Keywords

-

Funding

  1. Italian Ministry of Education, University and Research (MIUR)

Ask authors/readers for more resources

Balloon-expandable and self-expandable stents are the two types of coronary stents available. Basically, they differ in the modality of expansion. The present study analyses the stress state induced on the vascular wall, by the expansion of balloon-and self-expandable stents, using the finite element method. Indeed, modified mechanical stress state is in part responsible in the restenosis process. The balloon-expandable stents herein investigated are assumed to be made of stainless steel, while the self-expandable stents are made of a shape memory alloy. The effects of the severity of the coronary stenosis, the atherosclerotic plaque stiffness and the stent design are investigated. Comparing the self-expandable stent with the balloon-expandable one, the former induces fewer stresses and lower damage to the vessel, but, on the other hand, its lower stiffness induces a lower capability to restore vasal lumen and to contrast arterial elastic recoil.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available