4.6 Article

Influence of intra-fractional breathing movement in step-and-shoot IMRT

Journal

PHYSICS IN MEDICINE AND BIOLOGY
Volume 49, Issue 12, Pages N175-N179

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/49/12/N03

Keywords

-

Ask authors/readers for more resources

Efforts have been made to extend the application of intensity-modulated radiotherapy to a variety of organs. One of the unanswered questions is whether breathing-induced organ motion may lead to a relevant over-or underdosage, e.g., in treatment plans for the irradiation of lung cancer. Theoretical considerations have been made concerning the different kinds of IMRT but there is still a lack of experimental data. We examined 18 points in a fraction of a clinical treatment plan of a NSCLC delivered in static IMRT with a new phantom and nine ionization chambers. Measurements were performed at a speed of 12 and 16 breathing cycles per minute. The dose differences between static points and moving target points ranged between -2.4% and +5.5% (mean: +0.2%, median: -0.1%) when moving with 12 cycles min(-1) and between -3.6% and +5.0% (mean: -0.4%, median: -0.6%) when moving with 16 cycles min(-1). All differences of measurements with and without movements were below 5%, with one exception. In conclusion, our results underline that at least in static IMRT breathing effects (concerning target dose coverage) due to interplay effects between collimator leaf movement and target movement are of secondary importance and will not reduce the clinical value of IMRT in the step-and-shoot technique for irradiation of thoracic targets.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available