4.5 Article

Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 52, Issue 1, Pages 141-146

Publisher

WILEY
DOI: 10.1002/mrm.20110

Keywords

spin-lattice relaxation time; Look-Locker; heart; myocardium

Ask authors/readers for more resources

A novel pulse sequence scheme is presented that allows the measurement and mapping of myocardial T-1 in vivo on a 1.5 Tesla MR system within a single breath-hold. Two major modifications of conventional Look-Locker (LL) imaging are introduced: 1) selective data acquisition, and 2) merging of data from multiple LL experiments into one data set. Each modified LL inversion recovery (MOLLI) study consisted of three successive LL inversion recovery (IR) experiments with different inversion times. e acquired images in late diastole using a single-shot steady-state free-precession (SSFP) technique, combined with sensitivity encoding to achieve a data acquisition window of <200 ms duration. We calculated T-1 using signal intensities from regions of interest and pixel by pixel. T-1 accuracy at different heart rates derived from simulated ECG signals was tested in phantoms. T-1 estimates showed small systematic error for T-1 values from 191 to 1196 ms. In vivo T-1 mapping was performed in two healthy volunteers and in one patient with acute myocardial infarction before and after administration of Gd-DTPA. T-1 values for myocardium and noncardiac structures were in good agreement with values available from the literature. The region of infarction was clearly visualized. MOLLI provides high-resolution T-1 maps of human myocardium in native and post-contrast situations within a single breath-hold.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available