4.5 Article

Optimized conditions for diluting and reusing a fluorescent protein gel stain

Journal

ELECTROPHORESIS
Volume 25, Issue 15, Pages 2506-2510

Publisher

WILEY
DOI: 10.1002/elps.200406004

Keywords

fluorescent stain; protein detection; proteomics; SYPRO ruby

Funding

  1. NCI NIH HHS [R33 CA093292] Funding Source: Medline

Ask authors/readers for more resources

This study elucidates the optimum conditions at the minimum cost for using SYPRO Ruby protein gel stain. It deals with the effects of gel fixation and staining times, as well as dilution and reuse of SYPRO Ruby protein gel stain in one-dimensional (1-D) gels. Signal strength and dynamic range were highest in gels that were fixed thoroughly before staining, followed by overnight staining. Using the optimized protocol, dilution or reuse of the stain reduces the dynamic range and signal intensity. Sensitivity remains high if the stain is reused up to two times, but signal intensity is reduced up to 2.5-fold in twice used stain. Sensitivity also remains high if the stain is diluted 1:2 in water, but signal intensity is reduced up to 6-fold. Of the two options, reuse or dilution, reuse better retains signal intensity and dynamic range.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available