4.4 Article Proceedings Paper

Fast contact-mode atomic force microscopy on biological specimen by model-based control

Journal

ULTRAMICROSCOPY
Volume 100, Issue 3-4, Pages 253-257

Publisher

ELSEVIER
DOI: 10.1016/j.ultramic.2003.11.008

Keywords

atomic force microscopy

Categories

Ask authors/readers for more resources

The dynamic behavior of the piezoelectric tube scanner limits the imaging rate in atomic force microscopy (AFM). In order to compensate for the lateral dynamics of the scanning piezo a model based open-loop controller is implemented into a commercial AFM system. Additionally, our new control strategy employing a model-based two-degrees-of-freedom controller improves the performance in the vertical direction, which is important for high-speed topographical imaging. The combination of both controllers in lateral and vertical direction compensates the three-dimensional dynamics of the AFM system and reduces artifacts that are induced by the systems dynamic behavior at high scan rates. We demonstrate this improvement by comparing the performance of the model-based controlled AFM to the uncompensated and standard PI-controlled system when imaging pUC 18 plasmid DNA in air as well as in a liquid environment. (C) 2004 Elsevier B.V. All rights reserved.