4.4 Article

The design and performance of a portable handheld 11CO2 delivery system

Journal

APPLIED RADIATION AND ISOTOPES
Volume 94, Issue -, Pages 338-343

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2014.09.008

Keywords

[C-11]CO2; Radiochemistry; Portable; Miniature; PET

Funding

  1. U. S. Department of Energy, Office of Biological and Environmental Research [DE-AC02-98CH10886]

Ask authors/readers for more resources

We constructed a hand-held device to efficiently trap [C-11]CO2 from the cyclotron target, safely transport up to 3.7 GBq (100 mCi) doses to remote sites and release it without the need for a liquid cryogen. The system consists of a 180 W furnace and a miniature molecular sieve trap (80-100 mg; 80-100 mesh 13 x) placed inside a lead pig weighing 11.1 kg. The overall [C-11]CO2 delivery efficiency of the device is similar to 82% (> 99% trapping efficiency). Radiation dose rates measured at 30 cm from the surface of the pig are <43.5 mu Sv/h (5 mR/h) up to 2.59 GBq (70 mCi). (C) 2014 Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available