4.4 Article Proceedings Paper

PU and NP analysis of soil and sediment samples with ICP-MS

Journal

APPLIED RADIATION AND ISOTOPES
Volume 67, Issue 5, Pages 821-827

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2009.01.041

Keywords

Plutonium; Neptunium; Fusion; ICP-MS; Sediment; Soil

Ask authors/readers for more resources

A method to analyse Pu and Np was optimised to achieve low detection limits and high sample throughput. Soil and sediment samples were ashed and digested with a borate fusion. After dissolving the melt in nitric acid, Pu and Np were separated on a TEVA extraction chromatopraphy column. It was measured with a sector field ICP-MS. Detection limits in soils and sediments as low as 1 x 10(-15) g/g for Pu and Np were achieved. The method was applied to reference materials, soil profiles from Switzerland and sediment samples from the river Yenisei (RU), where radioactive nuclides have been discharged. (C) 2009 Elsevier Ltd. All rights reserved.