4.5 Article

Effect of lung mechanics on mechanically assisted flows and volumes

Journal

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/01.PHM.0000137309.34404.BC

Keywords

cough; assisted cough; mechanically assisted cough; exsufflation; respiratory care; noninvasive ventilation; neuromuscular disease

Ask authors/readers for more resources

Objective: To correlate the air flows generated by mechanical insufflation-exsufflation as a function of pressure delivery in a lung model at two pulmonary compliance and three airway resistance settings. Design: With each combination of pulmonary compliances of 25 and 50 ml/cm H2O and airway resistances of 6, 11, and 17 cm H2O/liter/sec, ten cycles of mechanical insufflation-exsufflation were applied using pressure deliveries of 40 to -40, 50 to -50, 60 to -60, and 70 to -70 cm H2O. The resulting peak exsufflation flows and volumes were recorded. Results: In a multivariate analysis, the pulmonary compliance, airway resistance, and pressure delivery were all found to significantly affect exsufflation flows and volumes such that a decreased pulmonary compliance or an increased airway resistance produced a decrease in exsufflation flow and volume, whereas an increased pressure delivery produced greater exsufflation flow and volume. Conclusion: Although mechanical insufflation-exsufflation pressures of 40 to -40 cm H2O are generally adequate for most patients with normal lung compliance and airway resistance, higher settings are often required when compliance decreases, by obesity or scoliosis, and possibly when airway resistance is increased.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available