In vitro and in vivo evaluation of a new active heat moisture exchanger

Journal

CRITICAL CARE
Volume 8, Issue 5, Pages R281-R288

Publisher

BIOMED CENTRAL LTD
DOI: 10.1186/cc2904

Keywords

absolute humidity; airflow resistance; heat moisture exchanger; hot water humidifiers; relative humidity

Ask authors/readers for more resources

Introduction In order to improve the efficiency of heat moisture exchangers (HMEs), new hybrid humidifiers (active HMEs) that add water and heat to HMEs have been developed. In this study we evaluated the efficiency, both in vitro and in vivo, of a new active HME (the Performer; StarMed, Mirandola, Italy) as compared with that of existing HMEs (Hygroster and Hygrobac; Mallinckrodt, Mirandola, Italy). Methods We tested the efficiency by measuring the temperature and absolute humidity (AH) in vitro using a test lung ventilated at three levels of minute ventilation (5, 10 and 15 l/min) and at two tidal volumes (0.5 and 1 l), and in vivo in 42 patients with acute lung injury (arterial oxygen tension/fractional inspired oxygen ratio 283 +/- 72 mmHg). We also evaluated the efficiency in vivo after 12 hours. Results In vitro, passive Performer and Hygrobac had higher airway temperature and AH (29.2 +/- 0.7degreesC and 29.2 +/- 0.5degreesC, [P < 0.05]; AH: 28.9 +/- 1.6 mgH(2)O/l and 28.1 +/- 0.8 mgH(2)O/l, [P < 0.05]) than did Hygroster (airway temperature: 28.1 +/- 0.3degreesC [P < 0.05]; AH: 27 +/- 1.2 mgH(2)O/l [P < 0.05]). Both devices suffered a loss of efficiency at the highest minute ventilation and tidal volume, and at the lowest minute ventilation. Active Performer had higher airway temperature and AH (31.9 +/- 0.3degreesC and 34.3 +/- 0.6 mgH(2)O/l; [P < 0.05]) than did Hygrobac and Hygroster, and was not influenced by minute ventilation or tidal volume. In vivo, the efficiency of passive Performer was similar to that of Hygrobac but better than Hygroster, whereas Active Performer was better than both. The active Performer exhibited good efficiency when used for up to 12 hours in vivo. Conclusion This study showed that active Performer may provide adequate conditioning of inspired gases, both as a passive and as an active device.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available